How to translate text using browser tools
23 September 2019 Determination of Fe, Ti in Seawater Using a Flow-injection Pretreatment System with ICP-MS
Huanying Ge, Ding Chen, Shu Li, Airong Zheng, Zuhao Zhu, Xuehong Zheng
Author Affiliations +
Abstract

Ge, H.; Chen, D.; Li, S.; Zheng, A.; Zhu, Z., and Zheng, X., 2019. Determination of Fe, Ti in seawater using a flow-injection pretreatment system with ICP-MS. In: Guido-Aldana, P.A. and Mulahasan, S. (eds.), Advances in Water Resources and Exploration. Journal of Coastal Research, Special Issue No. 93, pp. 77–82. Coconut Creek (Florida), ISSN 0749-0208.

In order to solve the interference caused by low metal concentration and salts matrix of seawater samples, this paper present a handy flow- injection system with solid-phase extraction to automate pretreatment procedures for offline Fe and Ti analysis in seawater by ICP-MS. By combining 8-position, 3-way and 6-way valves in the pipeline, the flow-injection system enhance the sample treatment efficiency by simultaneously loading two reagents online for: sample loading, buffering, and preconcentration. Each sample treatment spends about 16 min and consumes 12 mL of seawater sample and 8 mL of buffer solution. The chelation of Toyopearl-AF-Chelate-650M for Fe and Ti occurs in pH5.5 at an injection rate of 1.0 mL·min-1. The eluent (0.8 mol·L-1 HNO3) is able to elute Fe and Ti in 90 s at an elution rate of 1.0 mL·min-1. Seawater sample needs 90 min of 120 W UV-oxidation to dissociate metal-organic complexes. The precision and availability of this method are tested by measuring Jiulongjiang Estuary seawater samples.

©Coastal Education and Research Foundation, Inc. 2019
Huanying Ge, Ding Chen, Shu Li, Airong Zheng, Zuhao Zhu, and Xuehong Zheng "Determination of Fe, Ti in Seawater Using a Flow-injection Pretreatment System with ICP-MS," Journal of Coastal Research 93(sp1), 77-82, (23 September 2019). https://doi.org/10.2112/SI93-011.1
Received: 16 May 2018; Accepted: 5 June 2019; Published: 23 September 2019
KEYWORDS
Fe
flow-injection system
ICP-MS
Ti
trace metal